Application
Sheathed Thermocouple with high stability fixed-point cell; the fixed-point substance forces a typical signal during its melting and freezing (fixed-point plateau); calibration curve deviations of the whole measuring circuit will be
recognisable and could be corrected implementations with solid or adjustable thread joint are deliverable ( → chapter parts and accessories)
Bending radius of the sheath material≥ 5xød
Thermo junction
L (Fe-CuNi) DIN 43 710
J (Fe-CuNi) DIN EN 60584
K (NiCr-Ni) DIN EN 60584
N (NiCrSi-NiAl) DIN EN 60584
S (PtRh10-Pt) DIN EN 60584
B (PtRh30-PtRh6) DIN EN 60584
on inquiry
Number of Thermo junction
1 thermocouple
Fixed point material
Bi33In67 72,5 °C
In 156,59 °C
Sn 231,93 °C
Zn 419,53 °C
Al67Cu33 548,2 °C
Al87Si13 578,8 °C
Al 660,32 °C
Ag28Cu72 779,6 °C
on inquiry dependent on application temperature
Sheath diameter
1,5 mm
2,0 mm
3,0 mm
Sheath material
1.4541
1.4571
1.4841
2.4816
Measuring point design
insulated, integrated in fixed point cell
Nominal length L
on inquiry